Environment
High Volume Air Sampler Andersen Type HV-RW
HV-RW (for Andersen) is a particle separating air sampler, with an impactor equipped with a multi-stage porous jet nozzle. It is designed to collect particles suspended in the air that penetrate the respiratory system. The particle size separator main unit consists of 4 overlapping particle separating stages. It separates out aerosols between 1.1 and 7.0 μm in size. Aerosols smaller than 1.1 μm are filtered out by a backup filter.
The sampling flowrate is stabilized by combining a constant flowrate unit and a suction blower.
● Aerosols 1.1 to 7.0 μm in size are separated out by a 4-stage filter.
● Particles can be separated in 5 stages including the backup filter.
● The constant flowrate unit suppresses suction flowrate drops associated with collecting volume increases.
■ Applications
● Researching air pollution
● Researching respiratory disorders
● Analyzing carcinogenic substances separately by particle size
● In-factory monitoring
The sampling flowrate is stabilized by combining a constant flowrate unit and a suction blower.
● Aerosols 1.1 to 7.0 μm in size are separated out by a 4-stage filter.
● Particles can be separated in 5 stages including the backup filter.
● The constant flowrate unit suppresses suction flowrate drops associated with collecting volume increases.
■ Applications
● Researching air pollution
● Researching respiratory disorders
● Analyzing carcinogenic substances separately by particle size
● In-factory monitoring
Information by Item Code (Specifications)
Item Code | 080130-1209 | |
---|---|---|
Model | HV-RW(for Andersen) | |
Suction Pump | Brush-less blower | |
Flow Rate Detection | Differential pressure detection system | |
Power Failure Processing | Operation prior to the outage continues after recovery from the outage | |
Flow Accuracy | Within ±5 % of the set flowrate value | |
Particle Separation Range | 1.1 to 7.0μm | |
Particle Size Classification | 4 stages Stage 1:7.0μm and larger Stage 2:3.3 to 7.0μm Stage 3:2.0 to 3.3μm Stage 4:1.1 to 2.0μm (Backup filter:1.1μm and smaller) |
|
Mass | Approximately39kg | |
Suction Flow Rate | 566L/min |
|
Display | Touch panel LCD screen (with backlight) | |
Power Supply | AC100V 10A Cable length : 3m | |
Size | While using : 575(W)x575(D)x1420(H)mm, Collection port height : 1210mm(From the installation surface) Folded state : 575(W)x575(D)x960(H)mm |
|
Accessories | Dust separator main unit Collection paper (5 sheets each in 4 rows) 20 sheets Backup filters 5sheets |
|
Price (excluding tax) | Contact us | |