Specifications

General lab equipment

Waterbus TBM106AA type

* 1 Performance is when the room temperature is 20 ° C, a rip -shaped lid is placed with no samples.
* 2 The effective dimensions are the dimensions from the water surface 20 mm below the aquarium to the shelf board. The protrusion is not included.
* 3 Power plug is not included.
* 4 TBM212AA does not come with ground adapter.
* Do not use anything other than water for heat media.
* Use the water level in the aquarium to 20 mm or less from the top of the aquarium to prevent overflowing outside the aquarium, such as inflation of heat media due to the rise in temperature.

Information by Item Code (Specifications)

Item Code 046000-106
Model TBM106AA
Standard Inside dimensions in the tank [effective dimensions] * 2 830 (W) x 150 (d) x 100 (H) [45 (H)]
External dimensions 900 (W) x 290 (D) x 210 (H)
Con10ts about 12L
Number of lid holes 1 row 6 pieces
Lin diameter large diameter 115mm medium diameter 85mm small diameter 55mm
Power supply AC100V 14A
Mass about 14kg
Performance Used temperature range room temperature+5 ° C to boiling temperature
Temperature adjustment accuracy ± 3 ℃ AT 70 ° C
Temperature distribution accuracy ± 5 ° C AT 70 ° C
Temperature rising time Approximately 35 minutes room temperature → boiling temperature
Approximately tolerated surrounding temperature 5 ° C to 35 ° C
Composition Temperature control system liquid expansion type
Heater SUS316L Pipe heater
Heater capacity 1.4kw
Inner tank SUS304
Safety device empty heat prevention device, overcurrent / leakage breaker
Accessories 1 stainless steel lid, 1 type of wheel, 1 shelf plate, 2 pillars, Muff 1 type, 1 side bar, 1 type of clamp (0 ° C to 100 ° C) 1 piece, 1 ground adapter * 4
Price (excluding tax) Contact us
Note  

Related information

catalog