Autoclave Sterilier for lab Model SAC-1
              ・ The inside of the nozzle of the micropipet is also sterilized. Fully automatic from sterilization to drying.
・ Easy operation and daily sterilization work can be used with 100V power supply.
          ・ Easy operation and daily sterilization work can be used with 100V power supply.
 ・ The digital fit can be accommodated in 5 to 6 in the basket below. 
 ・ Rack chips, instruments, and plug -up containers can also be sterilized. (Sterilation of liquids such as medium is not possible.) 
 -The water for pressure is poured 2.8L from the top water supply. One consumption is small. 
Information by Item Code (Specifications)
| Item Code | 051800-400 | |
|---|---|---|
| Model | SAC-1 | |
| Safety features | Empty fire prevention device, safety valve, pressure safety switch, overcurrent protection breaker, do not operate when not locked door | |
| Sterilization temperature | 121 ° C/132 ° C. Switching | |
| Drying temperature | 90 to 100 ° C in the can  Can inner wall of can body about 160 ° C (can be changed to 90 minutes) Only drying can be operated There is a blaze function  | 
                |
| Volume / effective dimensions | 12.5L, φ220mm x 330mm | |
| Display method | Temperature / Time: Digital  Pressure: Analog  | 
                |
| Power Supply | AC100V 50/60Hz 1.44kw (the outlet uses 3 pintype) | |
| Size | 430 x 453 x 367mm 27kg | |
| Accessories | 1 snowboard plate, 1 drain filter, 1 nylon tashi, a sterile filter (attached to the intake port) | 
                |
| Price (excluding tax) | Contact us | |
| Note | ||