Specifications

Water Quality Test / Food Sanitary Control

Handy Water Meter AQUAB AQ-202

The AQUAB series handy water quality meters are easy-to-operate water quality photometers. As the concentration conversion curves are pre-programmed, measurements can be performed immediately after zero-point adjustment using sample water. Two-point calibration is enabled by setting to user calibration mode. Up to 99 measurement values can be stored in memory.

■ Main Functions
● Memory function
  Measurement values are automatically saved up to 99 points.
  In memory check mode, you can check the measurement value immediately.
● Turning the power OFF automatically or manually
  The power can be automatically turned OFF even if you forget to turn it OFF. You can also turn the power OFF manually.

● For measuring high-concentration residual chlorine
Sterile water such as a sodium hypochlorite solution and hypochlorous acid water (acidic electrolyzed water, etc.) are used in wide fields such as hygienic control of various types of foods including eggs and vegetables, towels and other items in hairdressing salons and barbers, and medical equipment as well as for the prevention of infection with diseases. To maintain a sufficient antiseptic effect, you must confirm that the residual chlorine concentration is surely maintained at a certain level. The DPD method, usually used as the method to measure residual chlorine concentration, has a characteristic that it cannot measure high-concentration residual chlorine.
The AQ-202 uses the iodine method and can measure high-concentration total residual chlorine up to 300 mg/L.

Information by Item Code (Specifications)

Item Code 080560-202
Model AQ-202
Measurement Principle Absorptiometry by Lodine reagent
Measurement Range 0~300mg/L
Light Source LED
Resolution 1mg/L / ±1mg/L
Measurement Item Residual chlorine (High concentration)
Measurement Method Permeation absorption measurement
Measurement Wavelength 511±2nm
Measuring Cell Vial bottle with 24mm cap
Display Format Digital Numbers
Data Recording 99points
Sample Water Volume 10mL
Automatic Power off Automatic power off after 120 seconds of key operation (manual operation is also possible)
Light Receiving Section Photodiode
Measurement Time Approx.2 seconds for both zero-point recognition and consentration measurement
Power Supply AAA batteries 4
Dimensions/Weight 70(W)x142(D)x63(H)mm - Approx. 250g
Accessories Powder reagents 25 packs
Sample cells 2 pieces
AAA alkaline batteries 4 pieces
Carrying case
Quick manual
Price (excluding tax) Contact us
Note  

Options and Spare Parts

Powder Reagent / high residual chlorine / 100 pks
AQUAB Test Tube / 10 mL / glassware