Specifications

Environment

Exhalation Valve Tester EXV-01

The Exhalation Valve Tester EXV-01 is designed to test according to JIS T 8151 the airtightness of exhaust valves used on particulate respirators. First, the tester draws air though an exhalation valve at a flow rate of 1 L/min and checks the degree of depressurization inside the valve caused by the exhaust valve closing airtight. Then,  the tester measures the time required for the internal pressure to reach atmospheric pressure.

The basic configuration consists of a pump to produce a depressurized state, a flowmeter for controlling the suction flow rate, a U-tube manometer for displaying pressure changes, 3 optical fiber sensors to detect the liquid levels in the manometer, and a timer to detect the time required for the liquid level to move between the optical fiber sensors.

Information by Item Code (Specifications)

Item Code ---
Model EXV-01
Timer 0.01 – 99.99 seconds
Suction Pump Flow rate: over 6.0 L/m
Suction pressure: over 0.016 MPa
Measurement Method Liquid level detection by fiber optic sensors and time measurement between sensors
Manometer U type glass tube, 0 – 2000 Pa
Flowmeter 0.2 - 2.0 L/min *equipped with a needle valve
Liquid level detection method Fiber optic sensors
Fiber optic sensor detection distance Maximum 50 mm
(standard sensing object: white paper)
Operating Environment 0 - 40℃, 0 - 90%RH *without condensation
Power Supply AC 100 V, 50/60 Hz. 1A
Dimensions/Weight W510 x D320 H360 (mm) / Approx. 14 kg
Price (excluding tax) Contact us